Model Number DSX520T11C
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2022
Event Type  malfunction  
Event Description
The manufacturer received information alleging the dreamstation 2 advanced auto cpap device had a scorched burn on the rubber pad on the bottom of the device, and it burned the nightstand.There was no report of patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for evaluation.Pil observed the rubber pad on the bottom of the device had dark markings on it.An unknown dust contaminant was observed on the ui bezel, top enclosure, center enclosure, iso port, bottom enclosure, blower, and blower outlet seal.The heater plate spring appeared to have a discoloration on it.The water tank was observed to have an unknown contaminant inside of it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION 2 ADV AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17753288
MDR Text Key323703365
Report Number2518422-2023-23087
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX520T11C
Device Catalogue NumberDSX520T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Initial Date Manufacturer Received 04/25/2022
Initial Date FDA Received09/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-