Catalog Number 6903270 |
Device Problems
Fluid/Blood Leak (1250); Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during an ultrasound guided dialysis catheter placement procedure via left femoral vein, the puncture needle was allegedly found to be broken which prevented the proper flow during puncture.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 11/2023).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that during an ultrasound guided dialysis catheter placement procedure via the left femoral vein, the puncture needle was allegedly found to be broken which prevented the proper flow during puncture.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Two electronic photos and one video were provided for review.The photos show one packed needle and packaging/labeling information, there is no visible fracture observed.In the video, the healthcare provider/staff tested the needle and it shows the needle unable to aspirate and leak is observed.Therefore, based on the photo and video review, the reported fracture could not be confirmed.The definitive root cause could not be determined based upon the available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2023).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|