• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 1000; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 1000; GENERATOR Back to Search Results
Model Number 1000
Device Problems Failure to Deliver Energy (1211); Low impedance (2285); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2022
Event Type  malfunction  
Event Description
It was reported that low impedance was seen with the patients generator.It was later reported that the patient received a battery replacement.During replacement procedure, lead was ensured to be properly secured and a test resistor was used and generator diagnostics was performed with an indication of low impedance.A new generator was implanted with patients current lead which resolved the high impedance.Device history records were reviewed for the generator.The device passed all functional specifications and quality tests and was hp sterilized prior to distribution.Programming history was received and reviewed confirming low impedance was seen on date of explant.No data prior to date of surgery was available.The cause of the low impedance event remains unknown.Explanted device has not been received to date.No other relevant information has been received to date.
 
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanovas employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any defects or malfunctions.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Explanted device has been received and is undergoing product analysis no other relevant information has been received to date.
 
Event Description
Product analysis was completed.Further assessment noted that the likely cause of the low impedance is due to a reed switch malfunction of being stuck close.No other relevant information has been received to date.
 
Event Description
Additional internal generator data was received form when low impedance was first observed.Analysis of this internal generator data provided an updated date of when the reed switch was stuck.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 1000
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key17759120
MDR Text Key323543286
Report Number1644487-2023-01318
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750405
UDI-Public05425025750405
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/06/2024
Device Model Number1000
Device Lot Number573
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received01/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-