• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE; RIGID SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE; RIGID SCOPE Back to Search Results
Model Number WA53005A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported that the scope had a dent.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: eyepiece damage.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The subject device was returned to an olympus repair center for evaluation and the customer¿s reported issue was confirmed.The device evaluation found that the lens was damaged.In addition, the device evaluation found that the lens inside was dirty and chipped.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to previous h6 (remove: 323 - gradient induced field problem).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the reportable malfunction occurred due to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
RIGID SCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17760595
MDR Text Key323827152
Report Number9610773-2023-02597
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052764
UDI-Public04042761052764
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53005A
Device Lot Number684844
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-