• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Device Problem Disconnection (1171)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 08/07/2023
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified access set spike separated from a viaflex bag containing norepinephrine during an infusion and medication leaked on the floor.The patient became hypotensive as a result.The nurse prepared another bag from the same lot and the separation occurred again.To resolve this issue, a bag from another lot was used and the patient¿s blood pressure recovered.No additional information is available.
 
Manufacturer Narrative
Upon follow up with the customer, it was reported that the separation occurred within the drug product and not between the bag and disposable set.It was also reported by the customer that the set that was used was a non-baxter product.Based on the additional information received, the baxter disposable was determined not to be a factor in the reported adverse event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE CORPORATION
ni
ni
ni
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17765396
MDR Text Key323621457
Report Number1416980-2023-04694
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
VIAFLEX IV
Patient Outcome(s) Other;
-
-