The reported field event of lead connection issue was confirmed.Septum material was observed inside the right ventricular lead set screws hex cavity.This did not allow the set screw to be tightened and secured properly in the field, resulting in high lead impedance.The issue was consistent with having occurred during the procedure.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
|