U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G247
Device Problems Over-Sensing (1438); High Capture Threshold (3266)
Patient Problems Chest Pain (1776); Dyspnea (1816)
Event Date 05/12/2020
Event Type  Injury  
Event Description
It was reported that while on a walk, the patient had chest tightness and a racing heartbeat.Two atrial tachycardia response (atr) episodes show an atrial rate of 120 beats-per-minute, but they were determined to be the result of far-field oversensing.Programming a blanking period was recommended to improve tracking.High thresholds were also noted.It was determined the events did not correlate with the patient symptoms.The patient was admitted to the hospital.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17770176
MDR Text Key323681051
Report Number2124215-2023-51012
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/01/2019
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number176048
Was Device Available for Evaluation? No
Date Manufacturer Received09/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age57 YR
Patient SexFemale
-
-