• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; NO MATCH Back to Search Results
Model Number CD3369-40Q
Device Problems Failure to Deliver Shock/Stimulation (1133); Failure to Convert Rhythm (1540); Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2023
Event Type  Injury  
Event Description
Related manufacture reference number: 2017865-2023-45400.It was reported that the patient admitted to the emergency room after receiving multiple high voltage shocks.Interrogation noted that the patient reentered ventricular tachycardia(vt) after anti-tachycardia pacing was provided.The second shock was not delivered due to low defibrillation lead impedance alert.The patient continued with vt storm and it was noted that the right ventricular lead also exhibited noise oversensing.The patient was successfully rescued.No interventions were performed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17770384
MDR Text Key323682876
Report Number2017865-2023-45364
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model NumberCD3369-40Q
Device Lot NumberA000085226
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/30/2023
Initial Date FDA Received09/18/2023
Supplement Dates Manufacturer Received01/24/2024
Supplement Dates FDA Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET LEAD; TENDRIL LEAD
Patient Outcome(s) Other;
Patient Age83 YR