Catalog Number 550310030 |
Device Problem
Osseointegration Problem (3003)
|
Patient Problems
Bone Fracture(s) (1870); Inadequate Osseointegration (2646)
|
Event Date 08/28/2023 |
Event Type
Injury
|
Event Description
|
It was reported that patient had reverse shoulder done on (b)(6) 2023.Surgeon informed that the patient had not been compliant in following post surgery restrictions.Patient continued doing activities which should not have done after surgery, which is believed to have lead to the glenosphere fracturing and becoming loose.Surgeon felt the bone quality that was left in the glenoid was not suitable for a revision and opted to convert the patient to a hemi shoulder.Surgeon removed most of the reverse implants, leaving the humeral component behind.A trial head reduction was performed and found to be adequate.The real head was opened and implanted.Doi: (b)(6) 2023.Dor: (b)(6) 2023.Affected side: right shoulder.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4) investigation summary no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|
|