• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIVIDEO CAMERA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIVIDEO CAMERA Back to Search Results
Model Number PVK-100S
Device Problem Optical Obstruction (3002)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2023
Event Type  malfunction  
Manufacturer Narrative
This device is not distributed in us so that 510k# is blank.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module with drive pcb.In addition, our technician confirmed that the root brace rubber (lg connector) cracked; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure(cloudy).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIVIDEO CAMERA
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key17778542
MDR Text Key323800897
Report Number9610877-2023-58993
Device Sequence Number1
Product Code FEA
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPVK-100S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/29/2023
Initial Date FDA Received09/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-