ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 146870489 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is not available for investigation.Without the return of the device a probable cause is unable to be determined.If additional information becomes available, a supplemental emdr will be submitted to the fda at that time.
|
|
Event Description
|
The complaint involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch that reportedly generated a proximal occlusion alarm on the plum 360 display.The b line infusion completed and the a line was started.They opened/closed pump, backprimed, and continued to rectify the alarm.A plum 360 infusion pump, ch 16 was used at the time the issue occurred.Although there was patient involvement, there was no human harm.There was an unspecified delay in patient therapy.
|
|
Manufacturer Narrative
|
The complaint of a proximal occlusion alarm was confirmed as a result of a lot history review.The probable cause of the occlusion alarm is related to material variability in cassette diaphragm stiffness which could cause an increase in the frequency of proximal occlusion alarms.The lot history was reviewed and the lot was found to fall under the scope of an existing corrective action that is in process.
|
|