Catalog Number 146870489
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2023
Event Type  malfunction  
Manufacturer Narrative
The device is not available for investigation.Without the return of the device a probable cause is unable to be determined.If additional information becomes available, a supplemental emdr will be submitted to the fda at that time.
 
Event Description
The complaint involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch that reportedly generated a proximal occlusion alarm on the plum 360 display.The b line infusion completed and the a line was started.They opened/closed pump, backprimed, and continued to rectify the alarm.A plum 360 infusion pump, ch 16 was used at the time the issue occurred.Although there was patient involvement, there was no human harm.There was an unspecified delay in patient therapy.
 
Manufacturer Narrative
The complaint of a proximal occlusion alarm was confirmed as a result of a lot history review.The probable cause of the occlusion alarm is related to material variability in cassette diaphragm stiffness which could cause an increase in the frequency of proximal occlusion alarms.The lot history was reviewed and the lot was found to fall under the scope of an existing corrective action that is in process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia IL
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key17779856
MDR Text Key323792831
Report Number9615050-2023-00457
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number146870489
Device Lot Number13518722
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PLUM 360 INF PUMP SN 20386835, MFR ICU MED
-
-