• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NCIRCLE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NCIRCLE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number N/A
Device Problems Loss of or Failure to Bond (1068); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Event Description
T was reported that a basket wire on a ncircle tipless stone extractor was found broken prior to use.The customer provided photo shows the basket wire detached from the basket sheath.The complaint device was not used on the patient, and a new same device was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1- customer (person): street: (b)(6).Line 2: (b)(6).Postal code: (b)(6).Phone: (b)(6).Mobile phone: (b)(6).E3- occupation: agent.G4- pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6 (annex a).Event summary.It was reported that a basket wire on a ncircle tipless stone extractor was found broken prior to use.The customer provided photo shows the basket wire detached from the basket sheath.The complaint device was not used on the patient, and a new same device was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Investigation evaluation.A document-based investigation was performed including a review of complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control procedures, as well as a visual inspection and functional test of the device were conducted.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.A database search did not identify any other events associated with the reported device lot.The information provided upon review of complaint file, device history record, complaint history, and quality control documents did not provide evidence to support that the device was manufactured out of specification or to suggest items in the lot or similar devices in the field or in house were nonconforming.Cook also reviewed product labeling.The product ifu, t ntse rev1; the ifu did not provide any information related to the reported issue.Based on the information provided, inspection of the returned device, and the results of the investigation, the cause could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCIRCLE TIPLESS STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key17783282
MDR Text Key323862038
Report Number1820334-2023-01267
Device Sequence Number1
Product Code FFL
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberNTSE-045065-UDH
Device Lot Number15379862
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received11/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight68 KG
-
-