MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  malfunction  
Event Description
A customer reported error message ¿2015 bf probe purge failure¿ during start-up on the aia-360 analyzer.The customer powered on/off the analyzer, but the error persists.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error log and attempted to perform bf prime, the error reoccurred.While troubleshooting, fse found defective bf syringe unit and 2-way solenoid valve (bf drain solenoid valve).Fse resolved the complaint by replacing the bf wash drain solenoid valve and bf syringe unit.Fse repaired and validated the analyzer by successfully performing daily check, and quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(6) from (b)(6) 2022 through aware date 21aug2023.There were no similar complaints identified during the search period.The aia-360 operator's manual under section7-1: error messages states the following: (2015) bf probe purge failure.Description: purging by the bf probe is abnormal.Troubleshooting: clean up the wash probe tip or replace it.Contact the service department.The most probable cause of the reported event was due to defective bf syringe unit and 2-way solenoid valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key17785213
MDR Text Key323907356
Report Number3004529019-2023-00419
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-