SYNTHES GMBH DEPTH GAUGE F/SCR Ø2+2.4 MEAS-RANGE UP-T Back to Search Results
Catalog Number 319.006
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported on an unknown date that some items were removed from a modular mini-fragment and needed to be replaced.2.0 and 2.4/2.7 bending irons are bleeding a rust-looking substance from the lettering printed.The drill guide cannot separate the inserts to disassemble because the threaded tip won¿t slide over the insert.The depth gauge tip is bent.The 2.7-module side is broken.There was no patient involvement, and no adverse events occurred.This report is for one (1)depth gauge f/scr ø2+2.4 meas-range up-t this is report 1 of 2 for complaint.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that depth gauge f/scr ø2+2.4 meas-range up-t was found with the needle deformed/bent.The protection sleeve was not received.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed deformed/bent condition of depth gauge f/scr ø2+2.4 meas-range up-t would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed the following source controlled drawings reflecting the current and manufactured revisions were reviewed: - depth gauge for 2.0 / 2.4mm screws 319_006_susa rev.N (current) / rev.E (manufactured).H4, h6 part # 319.006, synthes lot # 6525316, supplier lot #na, release to warehouse date: 17 nov 2010 , manufactured by: synthes jennersville.No non conformance reports were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE F/SCR Ø2+2.4 MEAS-RANGE UP-T
Type of Device
DEPTH GAUGE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17785362
MDR Text Key323877137
Report Number8030965-2023-11868
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number6525316
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/20/2023
Supplement Dates Manufacturer Received10/09/2023
Supplement Dates FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
2.7MM LCP(TM) MODULE CASE; 2.7MM UNIVERSAL DRILL GUIDE; BEND-PIN F/LCP-PL 2; BEND-PIN F/LCP-PL 2; BEND-PIN F/LCP-PL 2.4+2.7; BEND-PIN F/LCP-PL 2.4+2.7
-
-