Catalog Number HL-90 |
Device Problems
Device Alarm System (1012); False Negative Result (1225)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the disposable set alarm kept going off.Patient involvement unknown.
|
|
Manufacturer Narrative
|
Other, other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
Additional information received.The event occurred during biomedical engineering's acceptance testing prior to the device being put into service in the facility.No patient involvement.
|
|
Manufacturer Narrative
|
Email is:(b)(6).H3 and h6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection noted a missing lever on the micro switch.Functional testing found the disposable alarm went off continuously; confirming the customer complaint.Root cause was attributed to the missing lever on the micro switch; which triggers the device to circulate the water.The broken and missing microswitch lever was attributed to wear from usage.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the micro switch and o-rings.Performed preventative maintenance.Device passed functional testing after the completed repairs.
|
|