Model Number AIA-2000 ST
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2023
Event Type  malfunction  
Event Description
A customer reported error message "3005 leak sensor detected leakage" on the aia-2000 analyzer.The customer cleaned the wash probe, changed the tip and inspected for leakage with none identified.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp) and beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineer (fse) conducted a site visit and was able to confirm the problem by review of the error logs.During troubleshooting, the fse tested the bf washer performance to ensure no leaks were detected as well as visually inspecting the leak sensor, which was found to be covered in debris.The fse cleaned the leak sensor, ran a daily check, and quality control (qc) with results within acceptable range.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the search period.The aia-2000 operator's manual under appendix 4: error messages states the following: [3005] leak sensor detected leakage.Cause : the leak sensor under b/f unit detected leakage.Measurement is suspended.Solution : contact tosoh service center or local representatives.The most probable cause of the reported event was due to debris on the leak sensor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000 ST
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key17786304
MDR Text Key323896703
Report Number3004529019-2023-00421
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-