|
Model Number AIA-2000 ST |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/24/2023 |
Event Type
malfunction
|
Event Description
|
A customer reported error message "2235 b/f probe 1 suction failure" on the aia-2000 analyzer.The customer cleaned the wash probe and replaced the tip, but the analyzer will not complete a daily check or wash prime.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp) and beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Manufacturer Narrative
|
A field service engineer (fse) conducted a site visit and was able to confirm the problem by reviewing the error log as well as reproduce the problem by attempting to perform prime wash and the error re occurred.During troubleshooting the fse observed the b/f washing probe was unable to draw the wash properly, causing the fse to suspect an issue with the internal tubing.The fse cleaned the b/f probe 1 internal tubing with hot water and ten percent (10%) bleach.The fse validated the analyzer by running a daily check and quality control (qc) with results within acceptable range and no errors recurring.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the search period.The aia-2000 operator's manual under appendix 4: error messages states the following: [2235] b/f probe 1 suction failure cause : the overflow sensor detected liquid after washer suction action was completed.The measurement result will be flagged (wu flag).Solution : clean b/f probe 1.(see chapter 9, "maintenance") if retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event was due to obstruction of flow in the b/f washing probe internal tubing.
|
|
Search Alerts/Recalls
|
|
|