TOSOH HI-TEC, INC. AIA-2000 ST; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2023
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) conducted a site visit and was able to confirm the problem by reviewing the error log.Fse inspected the sorter and did not find any obstructions, then checked the tip alignment for every rack position in the sorter and needed to perform realignment.Fse performed tip alignments for every spot in the sorter for all the tip positions to correct the issue.Fse performed a sorter test for tip pick up with no issues after alignments were performed.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were two similar complaints identified during the search period, including this complaint.The aia-2000 operators manual under appendix 4: error messages state the following: [2207] no tip acquired by sorter cause : no tip was detected during the tip attachment check.If retry fails, measurement result will be flagged (mf flag).Solution : contact tosoh service center or local representatives.The most probable cause of the reported event was due to misaligned sorter tips.
 
Event Description
A customer reported error message "2207 no tip acquired by sorter¿ on the aia-2000 analyzer.The customer noted the first full box of tips are used with no errors, until there are ten (10) tips remaining, then the error occurs.The technical support specialist instructed the customer to check the waste chute and bin for any buildup.The customer called back at a later date and stated that the analyzer keeps skipping tests due to tip pickup error.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp) and beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000 ST
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key17786496
MDR Text Key323897521
Report Number3004529019-2023-00423
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-