Model Number AIA-2000 ST |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineer (fse) conducted a site visit and was able to confirm the problem by reviewing the error log.Fse inspected the sorter and did not find any obstructions, then checked the tip alignment for every rack position in the sorter and needed to perform realignment.Fse performed tip alignments for every spot in the sorter for all the tip positions to correct the issue.Fse performed a sorter test for tip pick up with no issues after alignments were performed.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were two similar complaints identified during the search period, including this complaint.The aia-2000 operators manual under appendix 4: error messages state the following: [2207] no tip acquired by sorter cause : no tip was detected during the tip attachment check.If retry fails, measurement result will be flagged (mf flag).Solution : contact tosoh service center or local representatives.The most probable cause of the reported event was due to misaligned sorter tips.
|
|
Event Description
|
A customer reported error message "2207 no tip acquired by sorter¿ on the aia-2000 analyzer.The customer noted the first full box of tips are used with no errors, until there are ten (10) tips remaining, then the error occurs.The technical support specialist instructed the customer to check the waste chute and bin for any buildup.The customer called back at a later date and stated that the analyzer keeps skipping tests due to tip pickup error.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp) and beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Search Alerts/Recalls
|
|
|