• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Break (1069); Unintended Power Up (1162); Appropriate Term/Code Not Available (3191); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2023
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: finland.The investigation is complete.This is an initial final report submission.Review of the most recent repair record determined the switch and machined head were damaged, the device was out of calibration, the motor was corroded, the motor speeds were unstable, and the motor function failed.The device is pending repair.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that before surgery the dermatome lever seems broken.Dermatome makes buzzing sound even when the safety is on.According to the customer, the dermatome did not work while pushing the lever.After that it seemed to be working even though no one applied pressure to the lever and at that time the safety was on.There was no patient involvement.No additional information available is indicated.Due diligence is complete.At product evaluation investigation the motor speeds were unstable.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17787710
MDR Text Key323908256
Report Number0001526350-2023-01175
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number61822722
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-