Product event summary: the afapro28 balloon catheter with lot number 07551 was returned and analyzed.No anomalies were identified during external visual inspection of the balloon, shaft, and handle segments.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for five applications on the reported event date.During functional testing, the console terminated the application and triggered system notice #12211 (the system did not recognize the catheter.) during inspection and pressure testing of the shaft segment, a guide wire lumen was partially detached from tip molding bond.During inspection of the handle segment, a broken thermocouple wire was identified.The thermocouple wire was broken/detached from the blue wire connection (connect to pin #two).In conclusion, the reported system notice 50005 was confirmed through testing and the balloon catheter failed the returned product inspection due to the guide wire lumen detached from the tip molding bond and the thermocouple wire broken inside the handle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|