Model Number 0998-00-3023-53
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cs300 intra-aortic balloon pump (iabp) unit failed battery run test.There was no patient involvement.
 
Manufacturer Narrative
The fse replaced the batteries to fix the issue.The fse then completed the pm with all functional and safety tests to factory specifications.The unit was returned to the customer.
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated data: b3,b4,g3,g6,h2,h10,h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17789817
MDR Text Key323943113
Report Number2249723-2023-04152
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment