BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G158 |
Device Problems
Premature Discharge of Battery (1057); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was beeping and recorded a code 1003, which is indicative of battery voltage being too low for the projected remaining capacity.Boston scientific technical services (ts) confirmed that the cause of the beeping was the 1003 message.Ts advised to send in data for analysis and explained how to turn off beeper.Device replacement was recommended.To date, this device remains in service.No adverse patient effects were reported.
|
|