Model Number 26605 |
Device Problems
Entrapment of Device (1212); Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that catheter entrapment occurred.The 50% stenosed target lesion was located in the mildly tortuous and non-calcified internal carotid artery.A 10.0-24 carotid wallstent was advanced for treatment.However, during the procedure the stent stuck on a non-boston scientific (bsc) wire and could not advance further.Hence, the device was removed by forcefully pulling out.Another stent was used, and the procedure was completed.There were no patient complications reported.
|
|
Event Description
|
It was reported that catheter entrapment occurred.The 50% stenosed target lesion was located in the mildly tortuous and non-calcified internal carotid artery.A 10.0-24 carotid wallstent was advanced for treatment.However, during the procedure the stent stuck on a non-boston scientific (bsc) wire and could not advance further.Hence, the device was removed by forcefully pulling out.Another stent was used, and the procedure was completed.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the carotid wallstent monorail 10.0-24 was returned for analysis.This carotid device is recommended for use with a 0.014" (0.36mm) guidewire as per instructions for use.During the product analysis a boston scientific 0.014" filterwire was successfully inserted through this device with no resistance experienced or lumen obstruction encountered.The guidewire used by the customer was not returned for analysis.A visual and tactile examination identified no damage or issues with the delivery system.The device was returned with the stent fully deployed from the device.No issues were noted with the deployed stent.
|
|
Search Alerts/Recalls
|