The olympus employee reported on behalf of the customer that during the right hemicolectomy procedure, short circuit error and probe tip damage occurred on thunderbeat 5 mm, 35 cm, front-actuated grip type s and the probe tip broke.The fallen device was immediately removed by the medical staff.The hospital had stock of a similar device, and the procedure was completed by replacing it with the new device and no delay was reported.There were no health hazards to the patient.
|
This report is being submitted to provide the legal manufacturer's final investigation results.Since the device was not returned to the legal manufacturer, olympus could not perform a physical evaluation.However, after reviewing a photo of the subject device, it was confirmed that the probe was broken.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, the root cause could not be determined.However, the broken probe was likely due to the following mechanisms: mechanism #1 1.During output activation in seal & cut mode, the probe came into contact with hard tissue, metal or surgical instruments.This caused scratches on the probe.This also caused short circuit error.2.A force to activate the output in seal & cut mode, or a force to grasp the body tissue was applied to the probe.Therefore, cracks were branching from the scratches on the probe.3.A force was applied to the probe causing it to break.Mechanism #2 1.Grasping section was closed without grasping anything while the device was activating in seal & cut mode (this includes after tissue resection) causing the tissue pad to wear out.2.Since the tissue pad was worn out, non-insulated area of the grasping section and the distal end of the probe came into contact.3.The output was activated in seal & cut mode in state of description stated above.This caused scratches (contact marks) on the distal end of the probe and the grasping section.The scratches indicate that the distal end of the probe was contacting the distal end of the grasping section.This also caused short circuit error.4.A force to activate the output in seal & cut mode, or a force to grasp the body tissue was applied to the probe.Therefore, cracks were branching from the scratches (contact marks) on the probe.5.A force was applied to the probe causing it to break.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.¿ ¿do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.¿ this supplemental report includes a correction to h3 from the initial medwatch.Olympus will continue to monitor field performance for this device.H3 other text : device was discarded at okr logistic site.
|