• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH UNKNOWN HEAD; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH UNKNOWN HEAD; HIP PROSTHESIS Back to Search Results
Catalog Number UNKNOWN
Device Problem Biocompatibility (2886)
Patient Problem Pain (1994)
Event Date 02/09/2022
Event Type  Injury  
Event Description
It was reported that approximately thirteen years post initial implantation, the patient underwent a left sided hip revision due to pain and elevated metal ions.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).G2 - foreign: italy.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2023-00526.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.H3 other text : product details unknown.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of manufacturing records cannot be performed without product identification.Device used for treatment.Insufficient information provided.Unable to perform a compatibility check.Complaint history review cannot be performed without product identification.Medical records were not provided.However, the legal claim received reference to a high level of ion metal: - plasma cobalt values: 30.63 g/l.- serum chromium values: 58.96 g/l.With the available information, a definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.Multiple supplemental mdr reports were filed for this event, please see associated reports: 0009613350-2023-00526-1.
 
Event Description
It was reported that approximately thirteen years post initial implantation, the patient underwent a left sided hip revision due to pain and elevated metal ions.Diligence was completed, no additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HEAD
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17797957
MDR Text Key324022010
Report Number0009613350-2023-00525
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/28/2023
Initial Date FDA Received09/22/2023
Supplement Dates Manufacturer Received11/02/2023
Supplement Dates FDA Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-