• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 42055150-120
Device Problem Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2023
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.A cine was received and reviewed by an abbott vascular clinical specialist.The complaint summary states that a 5.5 x 150 mm supera self-expanding stent was implanted in a superficial femoral artery.The summary states that upon implantation the stent was 50% longer than expected (calculated: 150 * 1.5 = 225 mm).There is a user technique to supera deployment which can cause the stent length to be longer or shorter than indicated.From the provided media (photos) there is nothing to suggest improper stent deployment from the positioning of the stent struts.Unfortunately, there is not a reference length in the media (such as a radiopaque ruler) so the length of the deployed stent cannot be confirmed.There is the possibility that a 180 or 200 mm length stent was incorrectly mounted in the system but the reported event cannot be confirmed with the information available.It is possible that during stent implantation interaction with the anatomy/other devices and/or pre-dilatation was not enough in some locations causing the stent to stretch/elongate after deployment; however, this could not be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that the procedure was to treat the superficial femoral artery without calcification or tortuosity.The 5.5x150mm supera self-expanding stent was implanted, however, it exceeded the length of the target lesion and was noted to have elongated 50% longer than expected.The stent remains in the patient anatomy.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17798354
MDR Text Key324031305
Report Number2024168-2023-10447
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number42055150-120
Device Lot Number2073061
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/02/2023
Initial Date FDA Received09/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
Patient Weight57 KG
-
-