Model Number CDVRA500Q |
Device Problems
Operating System Version or Upgrade Problem (2997); Wireless Communication Problem (3283)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a software upgrade anomaly occurred on the device.Further information was requested, but is not yet available.
|
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Manufacturer Narrative
|
Correction: upon review, the pulse generator should not have been submitted as a medical device report (mdr) as the event did not indicate a malfunction caused a serious event.
|
|
Event Description
|
Additional information received by abbott technical support indicated the device was in telemetry lockout.Software upgrade is not possible while the device in telemetry lockout.This is normal device behavior, and no device malfunction occurred.
|
|