• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) printer suddenly stopped working.Removing and reinserting the paper did not resolve the issue.This was the transport pump, so there was no other pump to switch to at that time.The pump has been requested to be sequestered and sent to biomed once the transport was completed.There was no patient harm reported.
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer(fse) was not dispatched as the issue was resolved over the phone with the customer.Service was handled by in-house biomed by flipping the printer paper, resolving the reported issue.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17801140
MDR Text Key324124299
Report Number2249723-2023-04184
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age81 YR
Patient SexPrefer Not To Disclose
Patient Weight96 KG
-
-