Model Number 0998-00-3023-53
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) battery run time failure.There was no patient involvement reported.
 
Manufacturer Narrative
Updated fields; b4, d9, g3, g6, h2, h3, h4, h6, h10.To fix the issue, the fse replaced the batteries.The fse then completed the pm with all functional and safety tests to factory specifications.The unit was returned to the customer.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17801860
MDR Text Key324129939
Report Number2249723-2023-04188
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received10/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2009
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-