• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83786
Device Problems Premature Activation (1484); Difficult to Remove (1528); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The main coil returned inside in a non-bsci device.Visual and microscope inspection revealed that the main coil was stretched, detached and bent at the coil arm section.No more damages can be observed on the device.The functional test could not be performed due only the main coil was returned for the analysis.Dimensional inspection in the main coil revealed that the zap tip od (outer diameter) and primary coil od were within specification.
 
Event Description
Reportable based on device analysis completed on 04sep2023.It was reported that the coil could not be advance halfway.The target lesion was located in the abdominal cavity.A 12mm x 20cm interlock-35 was selected for use.The device was continuously flush before introduction of the coil.During the procedure, it was noted that the coil could not be advance and withdrawn halfway even after several attempts.The whole catheter and the coil were pulled out together from the patient's body.It was then found that the coil had been detached inside the catheter.The procedure was completed with another of same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the main coil detached at the coil arm section.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Type of Device
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17801888
MDR Text Key324124982
Report Number2124215-2023-49595
Device Sequence Number1
Product Code KRD
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K133208
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number83786
Device Catalogue Number83786
Device Lot Number0030744019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/04/2023
Initial Date FDA Received09/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient Weight70 KG
-
-