• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134702
Device Problems Display or Visual Feedback Problem (1184); Material Puncture/Hole (1504); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and the bwi product analysis lab identified a hole on the pebax.During the procedure, when radiofrequency (rf) energy was applied to the smarttouch catheter the medical team lost visualization of the catheter and the force readings were inaccurate.Changing out the catheter cable did not resolve the issue.The catheter was changed and the issue resolved.The procedure was continued and subsequently completed.The inaccurate force reading is not mdr-reportable.Lost visualization is not mdr-reportable.Hole in the pebax is mdr-reportable.
 
Manufacturer Narrative
The product investigation was completed.Device evaluation details: the device was returned to biosense webster for evaluation.Visual inspection and functionality test of the returned device were performed following bwi procedures.Visual inspection was performed, and a hole was observed on the pebax surface with reddish material inside.The root cause of the damage on pebax cannot be related to the manufacturing process since there is evidence that the device was manufactured in accordance with documented specifications and procedures.However, it could be related to the handling of the device during the procedure, but this cannot be conclusively determined.Then, the catheter was connected to carto 3, and error 105 was observed due to an open circuit inside the tip.The customer complaint was confirmed.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: the magnetic sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.A manufacturing record evaluation was performed for the finished device batch number 31010124l, and no internal actions were identified.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17805394
MDR Text Key324121859
Report Number2029046-2023-02155
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 09/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134702
Device Lot Number31010124L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2023
Date Manufacturer Received08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-