Model Number 0998-00-0800-53
Device Problems Inaccurate Synchronization (1609); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cardiosave intra-aortic balloon pump (iabp) the stm found fault 124 & 58 in the log file, prolonged inflation with system shutdown.There was no patient involvement reported.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6, h10.The getinge service territory manager (stm) stated that the errors occurred on (b)(6) 2023 @ 22:58.The customer did not have any report of an incident from that date/time to report.The unit has run over 7,000,000 cycles since that date.The stm replaced the drive manifold and the drive transducer as a precautionary measure and completed a full pm.The stm ran the unit through the weekend, the unit was still running without error on monday.Returned to the customer and released for clinical use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17806392
MDR Text Key324116423
Report Number2249723-2023-04189
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-