It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cardiosave intra-aortic balloon pump (iabp) the stm found fault 124 & 58 in the log file, prolonged inflation with system shutdown.There was no patient involvement reported.
|
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6, h10.The getinge service territory manager (stm) stated that the errors occurred on (b)(6) 2023 @ 22:58.The customer did not have any report of an incident from that date/time to report.The unit has run over 7,000,000 cycles since that date.The stm replaced the drive manifold and the drive transducer as a precautionary measure and completed a full pm.The stm ran the unit through the weekend, the unit was still running without error on monday.Returned to the customer and released for clinical use.
|