Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).G2: foreign ¿ japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported by the surgeon that at an unknown point during a hip procedure, the tip of the shell inserter was broken, and the procedure was completed with another inserter.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the tip of the inserter to be fractured.A portion of the face of the tip remains attached.The fractured piece(s) were not returned.Scuffing and scratching are present on the shaft.The strike plate has been dinged.Scuffing is also present around the guide mating feature.The grip has become a dull, darker green color.Complaint confirmed based on evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17808305
MDR Text Key324146492
Report Number0001825034-2023-02237
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number724310
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-