Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).G2: foreign ¿ japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported by the surgeon that at an unknown point during a hip procedure, the tip of the shell inserter was broken, and the procedure was completed with another inserter.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the tip of the inserter to be fractured.A portion of the face of the tip remains attached.The fractured piece(s) were not returned.Scuffing and scratching are present on the shaft.The strike plate has been dinged.Scuffing is also present around the guide mating feature.The grip has become a dull, darker green color.Complaint confirmed based on evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|