Device evaluated by manufacturer: a synergy ous mr 3.00 x 12mm stent delivery system (sds) was returned for analysis.Visual, tactile, microscopic, dimensional and functional analysis was performed on the device.Visual examination identified stent damage along the entire length of the stent.A stent od reading could not be obtained.A review of the manufacturing stent profile data was performed and the stent od at the time of manufacture was 0.0403 inches which is within specification.Microscopic analysis of the stent identified stent struts pulled distally and bunched at the distal section.There were signs of movement, with the stent moved over the distal markerband.Balloon cones were reviewed, and no issues were noted.Balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.Bumper tip showed no signs of distal tip damage.No issues identified with the hypotube shaft.No issues identified with the outer / mid-shaft sections or the inner lumen of the device.The device was successfully tracked over a recommended 0.014 inch guidewire with no issues.
|