MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Deliver Energy (1211); Use of Device Problem (1670); Battery Problem (2885); Charging Problem (2892)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/18/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that information was received from a patient regarding an implantable neurostimulator for essential tremor and parkinsons.The reason for call was patient reported they were sent a new recharger, antenna and desktop charger 4 -5 days ago (see rtg0489427 and rtg0489447) and they plugged it in and it is not making it go and they are "starting to feel the effects", 2 days ago they noticed the felt like therapy may be off it and has been 4 days now since they were able to recharge.Pt said they recharge for 15 minutes to a half hour, every day to 100 percent.Worked with patient to use the equipment and patient reported seeing the reposition antenna screen.Patient repositioned the antenna several times but was not able to get the recharging screen.Asked pt to use their programmer to check their ins status and pt looked but stated they do not have a programmer they may have never had one.Reviewed information about discharged ins and without a programmer to check status and potentially turn therapy back on, they will likely need to see their doctor to start recharging again and turn therapy back on.Pt said they go to the va and does not remember their doctor's name.The issue was not resolved through troubleshooting.Offered and sent email to the reps in the area to get in touch with patient about getting their therapy turned back on and potentially help with a programmer or wireless recharger transition.Offered to replace the 37751 recharger through repairs but if the replacement does not show they are charging they will need to see their doctor or doctor/rep and they will need to see the doctor/ rep to turn therapy back on.Patient also reported 4 years ago they were switched to a rechargeable battery.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 37751 (serial: (b)(6)); product type: 0213-recharger; implant date n/a; explant date n/a.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the consumer reported the cause of the device being off and the recharging issue was the wire being frayed.The device was replaced which resolved the issue.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|