• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REP DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REP DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number UDSX500S11F
Device Problem Degraded (1153)
Patient Problems Pulmonary Edema (2020); Respiratory Tract Infection (2420); Balance Problems (4401); Unspecified Gastrointestinal Problem (4491)
Event Date 12/07/2022
Event Type  Injury  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged black specs blowing out of the device.The patient has also alleged of having lung infection, fluid in lungs, sinus infection, pressure in sinus, stomach problem and balance issue from the device.The patient consulted with lung doctor as well as pcp and was prescribed with medication for the infections.The device has been returned but not evaluated yet.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer previously reported receiving information in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged black specs blowing out of the device.The patient has also alleged of having lung infection, fluid in lungs, sinus infection, pressure in sinus, stomach problem and balance issue from the device.The patient consulted with lung doctor as well as pcp and was prescribed with medication for the infections.The device was returned to the manufacturer's product investigation laboratory for investigation and the customer's complaint was not confirmed.During the investigation, the manufacturer observed an unknown dust contaminant on the bottom enclosure and rear panel which is due to an outside environmental source.There was no evidence of foam degradation observed.Section d9, h3 and h6 were updated/corrected in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REP DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17812206
MDR Text Key324185448
Report Number2518422-2023-24375
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUDSX500S11F
Device Catalogue NumberUDSX500S11F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received09/25/2023
Supplement Dates Manufacturer Received02/29/2024
Supplement Dates FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-