• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; No Match Back to Search Results
Model Number 1458Q/86
Device Problems Separation Failure (2547); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Manufacturer Narrative
The reported events of stylet could not be removed from lead and connector pin detachment were confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found that the inner coil, connector pin and crimp sleeve were pulled out of the connector assembly, consistent with procedural damage.The ptfe coating of the stylet along with the inner coil was found to have bunched up/clogged distal to the connector pin.The cause of the reported events was due to the bunched up ptfe coating of the stylet inside the inner coil that prevented the removal of the stylet and excessive forces resulted in the connector pin and crimp sleeve to be pulled out of the connector assembly.
 
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, while unsuccessfully attempting to separate the stylet from the left ventricular (lv) lead, it was noted that the connector pin was detached from the body of the lead.The lv lead was removed and replaced during procedure.The patient was stable throughout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17814868
MDR Text Key324246197
Report Number2017865-2023-42534
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000142752
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/06/2023
Initial Date FDA Received09/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-