MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ PREFILLED NORMAL SALINE FLUSH SYRINGE; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ PREFILLED NORMAL SALINE FLUSH SYRINGE; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306546
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2023
Event Type  malfunction  
Manufacturer Narrative
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the plunger on the bd posiflush¿ prefilled normal saline flush syringe was difficult to move.Report 2 of 2.The following was received from the initial reporter: pharmacist called to report that there have been 2 occurrences of the plunger being difficult to move.1st on (b)(6) 2023 (lot# 3090870) and again on (b)(6) 2023 (lot# 3090870).There was no patient impact.
 
Event Description
It was reported that the plunger on the bd posiflush¿ prefilled normal saline flush syringe was difficult to move.Report 2 of 2.The following was received from the initial reporter: pharmacist called to report that there have been 2 occurrences of the plunger being difficult to move.1st on (b)(6) 2023 (lot# 3090870) and again on (b)(6) 2023 (lot# 3090870).There was no patient impact.
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 18-sep-2023.H6: investigation summary: it was reported the plunger is difficult to move.To aid in the investigation, two empty samples with no packaging flow wrap, or tip cap, and one photo was received for evaluation by our quality team.A visual inspection was performed, and no defects or imperfections were observed.Each sample was then tested for sustaining force, which is the force applied to the plunger rod while moving downwards when expelling the saline solution, and all results were within specification.The photo provided shows one of the samples received.A device history record review was completed for provided material number 306546, lot 3090870.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.To date, there have been no other similar events reported for this lot.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed and a probable root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ PREFILLED NORMAL SALINE FLUSH SYRINGE
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17815784
MDR Text Key324222926
Report Number1911916-2023-00686
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public(01)30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number306546
Device Lot Number3090870
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-