Catalog Number 03666794001 |
Device Problems
Display Difficult to Read (1181); Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is ongoing.
|
|
Event Description
|
We received an allegation of a display issue with coaguchek xs meter serial number (b)(6).The customer stated the meter display is faint and difficult to read.There was no misinterpretation of results.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.The returned meter did not power on.The meter was optically inspected and contamination from fluid ingress was observed on the printed circuit board.The contamination caused a power interruption which prevents the meter from powering on.Additionally, the contamination can cause faded/missing segments.The cause of the issue is contamination due to user error.
|
|