ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794001
Device Problems Display Difficult to Read (1181); Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.
 
Event Description
We received an allegation of a display issue with coaguchek xs meter serial number (b)(6).The customer stated the meter display is faint and difficult to read.There was no misinterpretation of results.
 
Manufacturer Narrative
The meter was returned for investigation.The returned meter did not power on.The meter was optically inspected and contamination from fluid ingress was observed on the printed circuit board.The contamination caused a power interruption which prevents the meter from powering on.Additionally, the contamination can cause faded/missing segments.The cause of the issue is contamination due to user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17817240
MDR Text Key324249396
Report Number1823260-2023-03102
Device Sequence Number1
Product Code GJS
UDI-Device Identifier04015630945689
UDI-Public04015630945689
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/26/2023
Supplement Dates Manufacturer Received10/04/2023
Supplement Dates FDA Received10/12/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-