Catalog Number 297500500
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "this instrument is composed of the 3 parts.The spring is missing from the reamer extension handle.As per the images attached." the product was not returned to depuy synthes, however photos were provided for review.See attachment 297500500.Jpg, 297500500 2nd.Jpg.The photo investigation revealed that the locking spring from reclaim dist reamer extension was missing.Review of provided photos shows that the locking spring is missing from the device, without having actual device for evaluation and without evaluating the spring component potential cause cannot be established.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the reclaim dist reamer extension would contribute to the complained device issue.Based on the investigation findings, potential cause cannot be established and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
This instrument is composed of the 3 parts.The spring is missing from the reamer extension handle.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.No further information can be obtained as the case was not reported by the customer.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: this instrument is composed of the 3 parts.The spring is missing from the reamer extension handle.As per the images attached.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed that there was no damage or defects with the reclaim dist reamer extension, the locking spring was on the device according to the drawing (dwg-297500500-component 2).The allegation cannot be confirmed.The overall complaint was unconfirmed as the observed condition of the reclaim dist reamer extension would not contribute to the complained device issue.Based on the investigation findings, the potential cause cannot be established, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DIST REAMER EXTENSION
Type of Device
RECLAIM INSTRUMENTS : ADAPTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17817675
MDR Text Key324257704
Report Number1818910-2023-19470
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295155911
UDI-Public10603295155911
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number297500500
Device Lot NumberNB76786
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received09/26/2023
Supplement Dates Manufacturer Received10/25/2023
12/26/2023
Supplement Dates FDA Received10/27/2023
12/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-