Model Number 0600-0783
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Event Description
A customer reported that during a patient procedure, using a glidescope core smart cable, the image was cutting in and out intermittently.No delay in the procedure, use of a backup device, or harm to the patient was reported.
 
Manufacturer Narrative
A replacement glidescope core smart cable was provided to the customer and the subject glidescope core smart cable used during the reported event was returned to verathon for evaluation.A verathon technical service representative evaluated the returned smart cable and was able to confirm the reported intermittent image issue.When connected to known, good, test verathon equipment, no image was produced and the image device was not recognized.The customer's smart cable failed verathon's device functionality testing.Upon completion of verathon's device evaluation, the smart cable was scrapped due to the customer already being provided a replacement and there being no repairs available for the device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key17818133
MDR Text Key324266397
Report Number9615393-2023-00188
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/08/2023
Initial Date FDA Received09/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-