Model Number 0600-0783 |
Device Problems
Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2023 |
Event Type
malfunction
|
Event Description
|
A customer reported that during a patient procedure, using a glidescope core smart cable, the image was cutting in and out intermittently.No delay in the procedure, use of a backup device, or harm to the patient was reported.
|
|
Manufacturer Narrative
|
A replacement glidescope core smart cable was provided to the customer and the subject glidescope core smart cable used during the reported event was returned to verathon for evaluation.A verathon technical service representative evaluated the returned smart cable and was able to confirm the reported intermittent image issue.When connected to known, good, test verathon equipment, no image was produced and the image device was not recognized.The customer's smart cable failed verathon's device functionality testing.Upon completion of verathon's device evaluation, the smart cable was scrapped due to the customer already being provided a replacement and there being no repairs available for the device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
|