Catalog Number 320559 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/11/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd nano¿ 2nd gen pen needle's cap would not attach.The following was translated from japanese to english: the customer reported that the pen attachment part was deformed.The pen attachment part was deformed and could not be attached.
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d9: device available for eval yes, d9: returned to manufacturer on: 11-oct-2023.H.6.Investigation summary : samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
|
Event Description
|
It was reported that the bd nano¿ 2nd gen pen needle's cap would not attach.The following was translated from japanese to english: the customer reported that the pen attachment part was deformed.The pen attachment part was deformed and could not be attached.
|
|