• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; No Match Back to Search Results
Model Number CDHFA500Q
Device Problems Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported that the patient presented for in-clinic follow-up.An interrogation of the implantable cardioverter defibrillator revealed that the device was in backup operation due to magnetic resonance imaging (mri) that was performed without appropriate programmed settings.High voltage therapy was disabled during backup operation.The device was successfully restored via programming.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17823273
MDR Text Key324328310
Report Number2017865-2023-47102
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberCDHFA500Q
Device Lot NumberS000078051
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/07/2023
Initial Date FDA Received09/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE; QUARTET; TENDRIL STS
Patient SexFemale
-
-