OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Defective Component (2292)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that there was a broken display.Patient involvement is unknown.
 
Manufacturer Narrative
Other, to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Manufacturer Narrative
Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Per visual inspection, scratches and nicks in the enclosure and front cover, pole clamp discolored, line cord faded, printed circuit board (pcb) had liquid crystal leakage, global display label had a dent where it covers the liquid crystal display (lcd), quick connect was corroded, water tank cover had scratches on it, missing reflux plug, interlock block contaminated.Visual inspection observed lcd to have liquid crystal leakage.The complaint was confirmed, and no other device analysis was performed.The most probable root cause was something impacted the front of the global display label enough to damage the lcd.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced pcb, membrane switch, quick connect, enclosure, water tank cover, line cord, pole clamp, interlock block, front cover, reflux plug.Performed preventative maintenance (pm) changed o-rings and reservoir gasket and calibrated.The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17827644
MDR Text Key324396418
Report Number3012307300-2023-09275
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer09/07/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/27/2023
Supplement Dates Manufacturer Received10/27/2023
Supplement Dates FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No