Model Number HL-90 |
Device Problems
No Audible Alarm (1019); Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a bench test, the device failed preventive maintenance.There was no alarm for the temperature being too high.No adverse effects, harm or injury occurred.
|
|
Manufacturer Narrative
|
Other, other text: d4: udi section is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Other text: d10, h3 and h6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found the device was received with front cover and enclosure having gouges in them.Quick connect was corroded.Printed circuit board (pcb) was missing an led.Pole clamp was discolored.Line cord faded.Water tank was contaminated with rust.Functional testing found the over "temp" did not alarm and the pump did not circulate.The complaint was confirmed.Root cause was attributed to a faulty printed circuit board (pcb).What caused this condition could not be established.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
|
|