• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Failure to Pump (1502); Failure to Calibrate (2440); No Flow (2991)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the pump was not functioning.It sounded like the pump was running, but there was no flow.The display showed that the temperature dropped and then quickly returned.They were unable to calibrate the display.Patient involvement is unknown.
 
Manufacturer Narrative
D4.Udi, d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device was received with the enclosure having some kind of black dye on the right side, front cover global display was dented, printed circuit board (pcb) missing a light emitting diode (led).Per functional testing, the pump was not circulating.The temperature would not rise.The display could not be calibrated, it stayed at 34.6.The complaint was confirmed.The root cause was a faulty pump and pcb.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17831500
MDR Text Key324707809
Report Number3012307300-2023-09335
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/06/2023
Initial Date FDA Received09/28/2023
Supplement Dates Manufacturer Received12/14/2023
Supplement Dates FDA Received01/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-