BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G125 |
Device Problems
Premature Discharge of Battery (1057); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the battery of this cardiac resynchronization therapy defibrillator (crt-d) was suspected to be depleting prematurely.The battery has three years remaining after being implanted five years ago.Boston scientific technical services (ts) discussed higher power consumption associated with sensing chronic atrial fibrillation (af).Ts also requested to have data from this device analyzed.Data analysis shows that the higher consumption is due to a high left ventricular (lv) threshold measurement leading to maximum voltage lv pacing.To date, this device remains in service.No adverse patient effects were reported.
|
|