Model Number 71992-01 |
Device Problem
Device Alarm System (1012)
|
Patient Problems
Hypoglycemia (1912); Coma (2417); Loss of consciousness (2418); Convulsion/Seizure (4406)
|
Event Date 09/04/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An alarm issue was reported with the adc device.The low glucose alarm did not sound, and customer was unaware of changes in glucose levels.As a result, customer experienced convulsion, epileptic seizure, loss of consciousness, and coma.Emergency service was called, and customer was transported to the hospital and received glucagon injection ¿via nasal way¿.Customer remained in the hospital for one week.No further treatment was required.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint.Dhrs for the fs libre reader and the dhrs showed the fs libre reader passed all tests prior to release.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An alarm issue was reported with the adc device.The low glucose alarm did not sound, and customer was unaware of changes in glucose levels.As a result, customer experienced convulsion, epileptic seizure, loss of consciousness, and coma.Emergency service was called, and customer was transported to the hospital and received glucagon injection ¿via nasal way¿.Customer remained in the hospital for one week.No further treatment was required.There was no report of death or permanent impairment associated with this event.
|
|