MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Display or Visual Feedback Problem (1184); Overheating of Device (1437); Use of Device Problem (1670); Battery Problem (2885); Charging Problem (2892); Communication or Transmission Problem (2896)
|
Patient Problem
Burning Sensation (2146)
|
Event Date 09/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: date is approximate.Year is confirmed valid.Section d references the main component of the system.Other medical products in use during the event include: brand name intellis; product id 97755 (serial: (b)(6)); product type: 0213-recharger medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient (pt) regarding an external device.It was reported that when charging the implantable neuros timulator (ins), it will have a lot of trouble finding the implant to charge and will say "charge controller battery" yet the controller is at 100 percent.This started about 3 weeks ago.They said they think their ins has "settled and sunk it to where its hard to know where it is".The recharger (rtm) is heating up and they felt like they were being burned".The issue was not resolved.A replacement recharger (rtm) was sent out.
|
|
Manufacturer Narrative
|
H3: product id 97755; serial# (b)(6).Was returned for product analysis.Analysis found that there was a failure with the recharger and that a "no device found" message was seen.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|