MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber is currently en route to fisher & paykel healthcare (f&p) in new zealand for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in china that a mr290v vented autofeed humidification chamber was found leaking water.There was no patient consequence.
 
Event Description
A distributor reported on behalf of a healthcare facility in china that a mr290v vented autofeed humidification chamber was found leaking water.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection of the returned mr290v vented autofeed humidification chamber confirmed multiple crack lines at the hinge bracket location near the base flange.Stress marks were found near the crack.Water was observed leaking from this crack.It was also noted that two vertical cracks above the water level were observed.Conclusion: we are unable to determine the cause of the reported event.However, our investigation indicates that the most likely cause is mechanical stress.The stress source is unknown.The mr290v chambers are designed and tested to conform to iso 5367 breathing tubes intended for use with anaesthetic apparatus and ventilators.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17840752
MDR Text Key324568615
Report Number9611451-2023-00876
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number2102323938
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/01/2023
Initial Date FDA Received09/28/2023
Supplement Dates Manufacturer Received10/26/2023
Supplement Dates FDA Received10/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P RT225 INFANT BIAS FLOW BREATHING CIRCUIT.; F&P RT225 INFANT BIAS FLOW BREATHING CIRCUIT.
-
-